NMR spectrometer Bruker Avance 600 is equipped with a 14 T superconducting magnet and two probes for 5 mm diameter liquid state samples: Double resonance broad band probe - BBO (an observe tunable multinuclear probe, optimized for X-nuclei detection) and Triple resonance broad band probe - TBI (an inverse tunable multinuclear probe, optimized for 1H detection). Broadband channel with frequency range 24 – 243 MHz can be optimized for nuclei from 107Ag to 31P. Both probes can use Z-gradient at all frequencies. Measurements can be performed in the temperature range of 223 - 373 K.