Principle of work
Thermogravimetry with model Mettler TG 50 is used for analysis the content of moisture and volatile components in the solid materials. The principle is based on monitoring of changes in the material during controlled heating. The analyses may be performed in a temperature range from room temperature to 900 ° C, in an air atmosphere. The sample is placed in a ceramic vessel that is connected to balance. It is heated within the furnace at a controlled rate and it's changes in weight was simultaneously recorded.